RVS-200 Integrated Modular Wall Diagnostic Station - Riester

RVS-200 Integrated Modular Wall Diagnostic Station

RVS-200 Integrated Modular Wall Diagnostic Station

RVS-200 is a versatile, integrated modular wall diagnostic station which like the RVS-100 measures pulse rate (PR), oxygen saturation (SpO2), non-invasive blood pressure (NIBP) and temperature (oral/axillary probe) with the additional option to add ophthalmoscopes, otoscopes and dermatoscopes. The flexible, modular design offers many configurations and settings to suit the needs of hospitals and clinics as well as office based and long-term care settings.



Integrated RVS-100 Vital Signs Monitor

Hardware

Features and benefits – hardware

Software

Features and benefits – software

Specifications: RVS-200

RVS-200

Power supply 100 – 240 V AC, 50/60 Hz
Battery Lithium-Ion battery, rechargeable, 10.8 V DC, 6,600 mAh. Charging time to a 100%: 6 hours, Operating time approx. 11 hours (new and fully charged battery at 25°C ambient temperature with connected SpO2, Temp, and NIBP on AUTO mode (15 minutes interval), Low battery indicator, Charging battery indicator
Monitor dimensions 303 x 230 x 110 mm (11.93 x 9.05 x 4.33 in)
Monitor weight 3600 g (7.94 lb) inc. temperature module, printer
Basket dimensions 270 x 225 x 200mm (10.63 x 8.86 x 7.87 in)
Alarm Visible LEDs (Yellow, Red, Blue, Green, Orange) Audible Speaker, Gives audible alarm, QRS tone. Supports Pitch Tone and multi-level volume. Alarm tones meet the requirement of IEC 60601-1-8. Alarm Pressure: 45 dB to 85 dB test distance is 1 meter from the tone
Ports 1 x AC power inlet
1 x standard RJ45 interfaces.100 BASE-TX, IEEE 802.3
4 x USB ports
1 x Equipotential grounding point
1 x RJ11 connector for nurse call
DC out port 15V/1.2 A

NIBP (non-invasive blood pressure)

Measuring method Automatic oscillometric
Measurement range Adult SYS 30 – 270 mmHg, DIA 10 – 220 mmHg, MAP 20 – 235 mmHg
Paediatric SYS 30 – 235 mmHg, DIA 10 – 220 mmHg, MAP 20 – 225 mmHg
Neonate SYS 30 – 135 mmHg, DIA 10 – 110 mmHg, MAP 20 – 125 mmHg
Cuff pressure range 0 – 280 mmHg
Resolution 1 mmHg
Pressure accuracy Static ±3 mmHg
Clinic ±5 mmHg, standard deviation ≤ 8 mmHg
Pulse rate range 40 – 240 bpm
Pulse rate accuracy 1 bpm
Initial inflation pressure Adult 160 mmHg, Paediatric 130 mmHg, Neonate 75 mmHg
Double overpressure protection (hardware and software).
Adult 297 ± 3 mmHg, Paediatric 252 ± 3 mmHg, Neonate 147 ± 3 mmHg
Alarm range SYS 0 – 300 mmHg
DIA 0 – 300 mmHg
MAP 0 – 300 mmHg

SpO2 – Riester

Measurement range 0 – 100%
Resolution 1%
Accuracy ±2% (70 to 100%)
±3% (40% to 69%)
Unspecified (0% to 39%)
Alarm range 0 – 100%
Average time 4 sec, 8 sec, 16 sec
Pulse rate
Measurement range
20 – 250 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy ±1% or ±1 bpm which ever is greater

SpO2 – Nellcor

Measurement range 0 – 100%
Resolution 1%
Accuracy 70 – 100%: ±2% (adult, paediatric)
70 – 100%: ±3% (neonate)
0 – 69%: unspecified
Alarm range 0 – 100%
Average time 8 sec, 16 sec
Pulse rate
Measurement range
20 – 300 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy 20 – 250 bpm: ±3 bpm
251 – 300 bpm: unspecified

SpO2 – Masimo

Measurement range 0 – 100%
Resolution 1%
Accuracy 70 – 100%: ±2% (adult, paediatric, non-motion conditions)
70 – 100%: ±3% (neonate, non-motion conditions)
70 – 100%: ±3% (motion conditions)
0 – 69%: unspecified
Alarm range 0 – 100%
Average time 2 – 4 sec, 4 – 6 sec, 8 sec, 10 sec, 12 sec, 14 sec, 16 sec
Pulse rate
Measurement range
25 – 240 bpm
Pulse rate Resolution 1 bpm
Pulse rate Accuracy ±3 bpm (non-motion conditions)
±5 bpm (motion conditions)

RPT-100 (Filac 3000) Predictive Thermometer

Oral (Quick Mode) non-fever temps: 3 - 5 seconds, fever temps: 8 - 10 seconds
Oral (Standard Mode) 6 - 10 seconds
Axillary Mode 8 – 12 seconds
Rectal Mode 10 – 14 seconds
Direct Mode (all sites) 60 – 120 seconds
Measurement accuracy 35.5°C - 42°C
Direct Mode (all sites): ±0.1°C (±0.2°F)
Standard Mode (all sites): ±0.1°C (±0.2°F)
Quick Mode (oral only): ± 0.3°C (±0.5°)

Inquiry about

Accept Terms and Conditions
RVS-200 Integrated Modular Wall Diagnostic Station
RVS-200 Integrated Modular Wall Diagnostic Station

You Recently Viewed Items

Events / Shows View All

AAO 2023
AAO 2023

AAO 2023

Bascom Palmer 2023
Bascom Palmer 2023

Bascom Palmer 2023

Reunión virtual con el Dr. Julian Triviño (Colombia)
Reunión virtual con el Dr. Julian Triviño (Colombia)

Reunión virtual con el Dr. Julian Triviño (Colombia)

AAO 2021
AAO 2021

 Friday, Nov. 12 - Monday, Nov. 15  Ernest N. Morial Convention CenterNew Orleans, LA.

5555 NW 74 AVE.
MIAMI. FL. 33166
P: 305-882-0120
F: 305-882-0140
sales@mercoframes.net
Skype mercoframesonline
Skype sales3mercoframes
Skype helpmercoframes
WhatsApp +1 305-882-0120
Hangouts ale@mercoframes.net